Description
FreeGuard® gloves are designed to help reduce exposure to scatter radiation while maintaining the comfort and dexterity needed for clinical procedures. The amount of radiation reduced, also called attenuation, depends on the glove thickness, X-ray tube voltage (kVp), and testing conditions. A higher attenuation percentage means a greater amount of radiation was reduced under the tested conditions.
- 0.22 mm thickness: Reduces radiation by 55% at 60 kVp, 44% at 90 kVp, and 37% at 120 kVp. Lead equivalency: 0.058 mm Pb.
- 0.30 mm thickness: Reduces radiation by 61% at 60 kVp, 49% at 90 kVp, and 42% at 120 kVp. Lead equivalency: 0.073 mm Pb.
What does lead equivalency mean?
Lead equivalency compares the glove’s radiation-reducing performance with the protection provided by a specified thickness of lead.
Testing was performed in accordance with IEC 61331-1:2014, with results provided by Emerson & Co. S.R.L.
Important: FreeGuard® gloves are designed to reduce exposure to scatter radiation. They are not intended for use in the primary X-ray beam.
FreeGuard® Regulatory Information: FDA 510(k) Clearance K192933